4-19
Performance Tests
7. Test Port Input Frequency Response
7. Test Port Input Frequency ResponsePerform this test to examine the vector sum of all test setup error vectors in 
both magnitude and phase change as a function of frequency. 
Required 
Equipment 
For 50Ω Analyzers 
Power Meter  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 436A/437B/438A
Power Sensor  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  Agilent 8482A
Cable, APC-7 24-inch  . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779
Adapter, APC-7 to Type-N (f) .  . . . . . . . . . . . . . . . . . . . . . . . . Agilent 11524A
Additional Equipment Required for Analyzers with Option 006 
Power Sensor  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  Agilent 8481A
For 75Ω Analyzers 
Power Meter  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 436A/437B/438A
Power Sensor  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent8483A Option H03
Cable, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  Agilent P/N 8120-2408
Procedure Power Meter Calibration for Test Port 1 from 300 kHz to 3 GHz 
1Zero and calibrate the power meter. 
2Connect the equipment as shown in Figure 4-10.