4-27
Performance Tests
7. Test Port Input Frequency Response
Test Port 1 Input Frequency Response from 3 GHz to 6 GHz 
39 When the analyzer finishes the calibration sweep, connect the equipment as 
shown in Figure4-15. 
Figure4-15. Setup for Test Port 1 Input Frequency Response 
40 Press MEAS, INPUT PORTS, A. 
41 Press MARKER, MARKER 1, MARKER FCTN, MARKER MODE MENU, MKR SEARCH, 
SEARCH:MIN to put marker 1 at the minimum magnitude location of the trace. 
42 Press MARKER, MARKER 2, MARKER FCTN, MARKER MODE MENU, MKR SEARCH, 
SEARCH:MAX to position marker 2 at the maximum magnitude location of the 
trace. 
43 Write the marker 1 or marker 2 reading, whichever has the largest absolute 
magnitude, in the “Performance Test Record.” 
Power Meter Calibration on Test Port 1 from 3 GHz to 6 GHz 
44 Connect the equipment as shown in Figure 4-16.