4-28
Performance Tests
7. Test Port Input Frequency Response
Figure4-16. Setup for Power Meter Calibration on Test Port 1 
45 Press MEAS, INPUT PORTS, TESTPORT 1. 
46 Press CAL, PWRMTR CAL, ONE SWEEP, TAKE CAL SWEEP to start the power meter 
calibration for output test port 1. 
Test Port 2 Input Frequency Response from 3 GHz to 6 GHz 
47 When the analyzer displays the message , POWER METER CALIBRATION 
SWEEP DONE, connect the equipment as shown as in Figure4-17.