6-23
Performance Tests Option 011
7. Receiver Magnitude Frequency Response
Input B Magnitude Frequency Response
13 Connect the equipment as shown in Figure 6-11.
Figure6-11. Magnitude Frequency Response Test Setup (Receiver Input B)
14 Press MEAS, INPUT PORTS, B, MENU to start measuring the power delivered to
receiver input B.
15 Press CW FREQ, 300, k/m.
16 Write the marker reading, which appears on the analyzer display, in the
B Input Power column of the Performance Test Record.
17 Repeat Step 15 and Step 16 for the other CW frequencies listed on the
Performance Test Record.
18 For each CW frequency, determine which channel input reading shows the
greatest difference:
power meter versus R
power meter versus A
power meter versus B
Refer to the first line for an example listed in the Performance Test Record .
Write the greatest difference in the space provided on the Performance Test
Record.