4-21
Performance Tests
7. Test Port Input Frequency Response
8Press MENU, NUMBER of POINTS, 51, x1.
9Press POWER, PWR RANGE MAN to turn the auto power range off.
10 Press PORT POWER to uncouple the test port output power.
11 Press CAL, PWRMTR CAL.
12 Press LOSS/SENSR LISTS, CAL FACTOR SENSOR A. Refer to the back of the power
sensor to locate the different calibration factor values along with their
corresponding frequencies.
The following softkeys are included in the sensor calibration factor entries
menu:
SEGMENT press to select a point where you can use the front panel
knob or entry keys to enter a value
EDIT press to edit or change a previously entered value
DELETE press to delete a point from the sensor calibration factor
table
ADD select this key to add a point into the sensor calibration
factor table
CLEAR LIST select this key to erase the entire sensor calibration factor
table
DONE select this key when done entering points to the sensor
calibration factor table
As an example, the following are the keystrokes for entering the first two cali-
bration factor data points for the Agilent 8482A power sensor (assuming
CF%= 96.4 at 100 kHz and CF%= 98.4 at 300 kHz):
NOTE
The analyzer displays the PRm annotation, indicating that the analyzer power range is
set to MANUAL.
NOTE
The analyzers calibration factor sensor table can hold a MAXIMUM of 12 calibration
factor data points.