4-21
Performance Tests
7. Test Port Input Frequency Response
8Press MENU, NUMBER of POINTS, 51, x1. 
9Press POWER, PWR RANGE MAN to turn the auto power range off. 
10 Press PORT POWER to uncouple the test port output power. 
11 Press CAL, PWRMTR CAL. 
12 Press LOSS/SENSR LISTS, CAL FACTOR SENSOR A. Refer to the back of the power 
sensor to locate the different calibration factor values along with their 
corresponding frequencies. 
The following softkeys are included in the sensor calibration factor entries 
menu: 
SEGMENT   press to select a point where you can use the front panel 
knob or entry keys to enter a value 
EDIT   press to edit or change a previously entered value 
DELETE   press to delete a point from the sensor calibration factor 
table 
ADD   select this key to add a point into the sensor calibration 
factor table 
CLEAR LIST   select this key to erase the entire sensor calibration factor 
table 
DONE   select this key when done entering points to the sensor 
calibration factor table 
As an example, the following are the keystrokes for entering the first two cali-
bration factor data points for the Agilent 8482A power sensor (assuming 
CF%= 96.4 at 100 kHz and CF%= 98.4 at 300 kHz): 
NOTE 
The analyzer displays the PRm annotation, indicating that the analyzer power range is 
set to MANUAL. 
NOTE
The analyzer’s calibration factor sensor table can hold a MAXIMUM of 12 calibration 
factor data points.