4-60
Performance Tests
16. Test Port Receiver Phase Compression
For 75Ω Analyzers 
Cable, 75Ω, Type-N, 24-inch . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-2408
Procedure Test Port 2 Phase Compression 
1Connect the equipment as shown in Figure 4-35. 
Figure4-35. Test Port Phase Compression Test Setup 
2Press PRESET, MEAS, Trans: E/E S21 FWD, FORMAT, PHASE. 
3Press AVG, IF BW, 10, x1. 
4Press MENU, SWEEP TYPE MENU, POWER SWEEP, START, –10, x1. 
5Press MENU, CW FREQ, 50, M/µ. 
6Press MENU, TRIGGER MENU, SINGLE. 
7At the end of the sweep, press SCALE REF, AUTO SCALE. 
8Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH: MAX. 
9Press MARKER, MARKER 2, MARKER FCTN, MARKER MODE MENU, MKR SEARCH, 
SEARCH: MIN. 
10 Press MARKER, ∆MODE MENU, ∆REF= 1. 
11 Write the absolute value of the marker 2 reading in the “Measured Value” 
column of the “Performance Test Record.” 
12 Repeat Step 5 through Step 11 for the other CW frequencies listed for Port 2  in 
the “Performance Test Record.”