5-11
Automated Verification Option 011
In Case of Difficulty
8When the analyzer finishes all the measurements, reconnect the mismatch
verification device, as shown in Figure 5-7.
Figure5-7. Mismatch Device Verification Setup
9Press RETURN, TESTS, 31, x1, EXECUTE TEST, CONTINUE.
10 You have completed the system verification procedure when the analyzer
displays, TEST 31 Ver Dev 4 PASS.
In Case of Difficulty
1Inspect all connections. Do not disconnect the cables from the analyzer test
ports. Doing so will invalidate the calibration that you have done earlier.
2Press PRESET, SAVE/RECALL.
aUsing the front panel knob, highlight the title of the full 2-port measurement
calibration that you have done earlier.
bPress RECALL STATE.
cRepeat Step 3. Device Verification on page 5-8.