5-17
Automated Verification – Option 011
Step 3. Device Verification
as shown in Figure5-11. 
Figure5-11. Connections for the 20 dB Attenuator 
3Press CONTINUE to run the test: 
•If you switched the record function off, you have to press CONTINUE after 
each S-parameter measurement. 
•If you switched the record function on, the test displays and prints the pass/
fail information for the S-parameter measurements that are valid for system 
verification.