3-5
Manual Verification
Step 3. Device Verification
18 Connect the test port cables together as shown in Figure 3-2.
Figure3-2. Transmission Calibration Setup
19 Press TRANSMISSION, FWD TRANS THRU.
20 When the Agilent8702D finishes measuring the standard, press FWD MATCH
THRU.
21 When the Agilent8702D finishes measuring the standard, press REV TRANS
THRU.
22 When the Agilent8702D finishes measuring the standard, press REV MATCH
THRU.
23 When the Agilent8702D finishes measuring the standard, press STANDARDS
DONE, DONE 2-PORT CAL.
24 Press SAVE/RECALL, SELECT DISK, INTERNAL MEMORY, RETURN, SAVE STATE to save
the measurement calibration into the Agilent 8702D internal mem ory.
25 Press SELECT DISK, INTERNAL DISK.
Step 3. Device Verification
1Press SYSTEM, SERVICE MENU, TESTS, 28, x1, EXECUTE TEST.
2Connect the 20 dB attenuator (supplied in the verification kit ) as shown in