
6-5
Performance Tests – Option 011
2. Source Power Range, Linearity, and Accuracy
2. Source Power Range, Linearity, and AccuracyPerform this test to verify the analyzer’s power range and linearity at different
CW frequencies throughout the analyzer operating frequency rang e.
Required
Equipment
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 436A/437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8482A
2-way Power Splitter (2), 50Ω . . . . . . . . . . . . . . . .Agilent11667A Option 001
Attenuator (2), 20 dB . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8491A Option 020
RF Cable Set, 50Ω, Type-N. . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 11851B
Additional Required Equipment for Analyzers with Option 006
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8481A
Procedure Path loss calibration
1Connect the equipment as shown in Figure 6-2.
Figure6-2. Path Loss Calibration Test Setup (#1)