6-41
Performance Tests Option 011
12. Receiver Magnitude Dynamic Accuracy
cPress AVG, IF BW, 10, x1.
4Connect the equipment as shown in Figure 6-22.
Figure6-22. Channel A Magnitude Dynamic Accuracy Test Setup
5On the Agilent 8702D Option 011:
aPress MEAS, INPUT PORTS, A, MENU, POWER.
bUsing the analyzer front panel knob, adjust the test port power for a reading
of 30.000 dBm4 on the Agilent 8902A.
cWait for the reading to stabilize.