6-41
Performance Tests – Option 011
12. Receiver Magnitude Dynamic Accuracy
cPress AVG, IF BW, 10, x1. 
4Connect the equipment as shown in Figure 6-22. 
Figure6-22. Channel A Magnitude Dynamic Accuracy Test Setup 
5On the Agilent 8702D Option 011: 
aPress MEAS, INPUT PORTS, A, MENU, POWER. 
bUsing the analyzer front panel knob, adjust the test port power for a reading 
of –30.000 dBm4 on the Agilent 8902A. 
cWait for the reading to stabilize.