5-2
Automated Verification Option 011
Automated Performance Verification for Option 011
Automated Performance Verification for Option 011
The automated performance verification procedures in this chapter provide a
high level of confidence that the Agilent8702D Option 011 is working prop-
erly. Performance verification procedures are provided for the Agilent8702D
Option 011 using the:
1Agilent 85046A or 85047A S-Parameter Test Set
2Agilent 85044A Transmission/Reflection Test Set
Although these procedures are automated, they do not require a computer.
However, youll need to locate the Verification Data Disk.
Allow the Agilent 8702D Option 011 to warm up for one hour before starting
this procedure.
Agilent Technologies recommends that you verify your analyzer measurement
system every six months. Agilent Technologies also suggests that you get your
verification kit recertified annually. Refer to the Agilent 85029B Option 001
7 mm Verification Kit Operating and Service Manual for more informa-
tion.
Agilent8753D is shown in the figures
The procedures in this chapter are designed to work with Agilent 8702D and
Agilent8753D instruments. Because of thi s, many of the figures label the instrument
being tested as an Agilent8753D. Simply substitute your Agilent 8702D for the
Agilent8753D.