6-39
Performance Tests – Option 011
12. Receiver Magnitude Dynamic Accuracy
20 Repeat Step 7 through Step 12 to measure the receiver R input impedance.
In Case of 
Difficulty 
WARNING These servicing instructions are for use by qualified personnel only. 
To avoid electrical shock, do not perform any servicing unless you  are 
qualified to do so.
WARNING The opening of covers or removal of parts is likely to expose 
dangerous voltages. Disconnect the instrument from all voltage 
sources while it is being opened.
WARNING The power cord is connected to i nternal capacitors that may remain 
live for five seconds after disconnecting the plug from its power 
supply.
1Check your test setup for damage or excessive wear of the i nput connectors.
2Impedance mismatch may also be caused by the samplers, or the semi-rigid 
cables connecting the A4/A5/A6 sampler mixer assemblies to the analyzer front 
panel Type-N connectors. Check for damaged semi-rigid cables. If no damage 
is apparent at the input connectors, try interchanging the sampler assemblies. 
If the problem goes away, the sampler which you just swapped is defective. If 
the problem still exists, suspect the path from the Type-N connector input to 
the sampler assembly.
3Refer to the Agilent8753 D Option 011 Service Guide for more 
troubleshooting information. 
12. Receiver Magnitude Dynamic AccuracyPerform this procedure to measure the magnitude dynamic accuracy for each 
input test port.