4-40
Performance Tests
10. System Trace Noise (Only for Analyzers without Option 006)
Source Match (Reverse) Calibration Coefficient 
69 Press SYSTEM, SERVICE MENU, TESTS, 39, x1, EXECUTE TEST. 
70 When the analyzer finishes the test, repeat  Step 63 through Step 66. 
Reflection Tracking (Reverse) Calibration Coefficient 
71 Press SYSTEM, SERVICE MENU, TESTS, 40, x1, EXECUTE TEST. 
72 When the analyzer finishes the test, repeat  Step 63 through Step 66. 
10. System Trace Noise (Only for Analyzers without Option 006)Perform this test to measure the system trace noise at a designated frequency 
in both the A/R and B/R ratioed measurements. 
Required 
Equipment 
For 50Ω Analyzers 
Cable, APC-7, 24-inch  . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent P/N 8120- 4779
For 75Ω Analyzers 
Cable, 75Ω, Type-N 24-inch  . . . . . . . . . . . . . . . . . . . . .Ag ilent P/N 8120-2408
Procedure
1Connect the equipment as shown in Figure 4-24. 
Figure4-24. System Trace Noise Test Setup