4-40
Performance Tests
10. System Trace Noise (Only for Analyzers without Option 006)
Source Match (Reverse) Calibration Coefficient
69 Press SYSTEM, SERVICE MENU, TESTS, 39, x1, EXECUTE TEST.
70 When the analyzer finishes the test, repeat Step 63 through Step 66.
Reflection Tracking (Reverse) Calibration Coefficient
71 Press SYSTEM, SERVICE MENU, TESTS, 40, x1, EXECUTE TEST.
72 When the analyzer finishes the test, repeat Step 63 through Step 66.
10. System Trace Noise (Only for Analyzers without Option 006)
Perform this test to measure the system trace noise at a designated frequency
in both the A/R and B/R ratioed measurements.
Required
Equipment
For 50 Analyzers
Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent P/N 8120- 4779
For 75 Analyzers
Cable, 75, Type-N 24-inch . . . . . . . . . . . . . . . . . . . . .Ag ilent P/N 8120-2408
Procedure
1Connect the equipment as shown in Figure 4-24.
Figure4-24. System Trace Noise Test Setup