2-5
Automated Verification
Step 2. Measurement Calibration
Figure2-2. Connections for Measurement Calibration Standards
5Press FORWARD:OPEN.
6When the Agilent 8702D finishes measuring the standard, connect the short
end of the open/short combination to reference test port 1.
7Press FORWARD:SHORT.
8When the Agilent8702D finishes measuring the standard, connect the 50 ohm
termination (supplied in the calibration kit) to reference test port 1.
9Press FORWARD:LOAD.
10 When the Agilent8702D finishes measuring the standard, connect the open
end of the open/short combination to reference test port 2.
11 Press REVERSE:OPEN.
12 When the Agilent 8702D finishes measuring the standard, connect the short
end of the open/short combination to reference test port 2.
13 Press REVERSE:SHORT.
14 When the Agilent8702D finishes measuring the standard, connect the 50 ohm
termination to reference test port 2.
15 Press REVERSE:LOAD.
16 When the Agilent8702D finishes measuring the standard, press STANDARDS
DONE. The Agilent 8702D briefly displays, COMPUTING CAL COEFFICIENTS.