4-39
Performance Tests
9. Calibration Coefficients
Figure4-23. Transmission Calibration Test Setup
57 Press TRANSMISSION, FWD TRANS THRU.
58 When the analyzer finishes the measurement , press FWD MATCH THRU.
59 When the analyzer finishes the measurement , press REV TRANS THRU.
60 When the analyzer finishes the measurement , press REV MATCH THRU.
61 When the analyzer displays, PRESS DONE IF FINISHED WITH STD(s),
press STANDARDS DONE, DONE 2-PORT CAL.
Load Match (Forward) Calibration Coefficient
62 Press SYSTEM, SERVICE MENU, TESTS, 36, x1, EXECUTE TEST.
63 When the test is done, press MARKER, MARKER 1.
64 Using the front panel knob, locate the maximum value of the data trace for the
300 kHz to 1.3 GHz frequency range.
65 Write the maximum value on the Performance Test Record.
66 Repeat the previous three steps for the other frequency range(s) listed on the
Performance Test Record.
Directivity (Reverse) Calibration Coefficient
67 Press SYSTEM, SERVICE MENU, TESTS, 38, x1, EXECUTE TEST.
68 When the analyzer finishes the test, repeat Step 63 through Step 66.