4-77
Performance Tests
Performance Test Records
Test 6. Test Port Input Noise Floor Level
Frequency
Range Test Port IF
Bandwidth
Specification
(dBm)
Calculated
Value
Measurement
Uncertainty
50 kHz 3 GHz Port 1 3 kHz 82 ____________ N/A
50 kHz 3 GHz Port 1 10 Hz 102 ____________ N/A
50 kHz 3 GHz Port 2 10 Hz 102 ____________ N/A
50 kHz 3 GHz Port 2 3 kHz 82 ____________ N/A
Option 006
3 GHz 6 GHz Port 2 3 kHz 77 ____________ N/A
3 GHz 6 GHz Port 2 10 Hz 97 ____________ N/A
3 GHz 6 GHz Port 1 10 Hz 97 ____________ N/A
3 GHz 6 GHz Port 1 3 kHz 77 ____________ N/A
Test 7. Test Port Input Frequency Response
Frequency
Range Test Port Specification
(dB)
Measured
Value (dB)
Measurement
Uncertainty (dB)
300 kHz 3 GHz Port 2 ± 1 _________________ 0.15
300 kHz 3 GHz Port 1 ± 1 _________________ 0.15
Option 006
3 GHz 6 GHz Port 1 ± 2 _________________ 0.18
3 GHz 6 GHz Port 2 ± 2 _________________ 0.18