6-52
Performance Tests Option 011
14. Receiver Magnitude Compression
20 Press STOP and enter the Stop Power value as determined in Step 18.
21 Press MENU, TRIGGER MENU, SINGLE.
22 At the end of the sweep, press SCALE REF, AUTO SCALE.
23 Press MARKER, MODE MENU, REF=1, MARKER FCTN, MARKER MODE MENU, MKR
SEARCH, SEARCH: MAX.
24 Press MARKER, MARKER 2, MARKER FCTN, MARKER MODE MENU, MKR SEARCH,
SEARCH: MIN.
25 Write the absolute value of marker 2 in the Measured Value column on the
"Performance Test Record."
26 Repeat Step 15 through Step 25 for the other frequencies listed on the
Performance Test Record.