5-6
Automated Verification – Option 011
Step 2. Measurement Calibration
Step 2. Measurement Calibration 
1Press CAL, CAL KIT & STDS, SELECT CAL KIT, CAL KIT:7mm, RETURN, RETURN, CALIBRATE 
MENU, FULL 2-PORT. 
2Press ISOLATION, OMIT ISOLATION, ISOLATION DONE. 
3Press REFLECTION. 
4Connect the “open” end of the open/short combination (supplied in the 
calibration kit) to reference test port 1, as shown in Figure5-2. 
Figure5-2. Connections for Measurement Calibration Standards 
5Press FORWARD:OPEN. The analyzer displays the message WAIT--MEASURING 
CAL STD.
6When the analyzer finishes measuring the standard, connect the “short” end of 
the open/short combination to reference test port 1, as shown in Figure5-2.