4-36
Performance Tests
9. Calibration Coefficients
Figure4-21. Transmission Calibration Test Setup 
19 Press TRANSMISSION, FWD TRANS THRU. 
20 When the analyzer finishes the measurement , press FWD MATCH THRU. 
21 When the analyzer finishes the measurement , press REV TRANS THRU. 
22 After the analyzer finishes the measurement, press REV MATCH THRU. 
23 After the analyzer finishes the measurement, press STANDARDS DONE, DONE 2-
PORT CAL. 
Directivity (Forward) Calibration Coefficient 
24 Press SYSTEM, SERVICE MENU, TESTS, 32, x1, EXECUTE TEST. 
25 When the analyzer finishes the test, press  MARKER. 
26 Using the front panel knob, locate the maximum value of the data trace for the 
300 kHz to 1.3 GHz frequency range. 
27 Write the maximum value in the “Performance Test Record.” 
28 Repeat the previous two steps for the other frequency range(s) listed on the 
“Performance Test Record.” 
Source Match (Forward) Calibration Coeffic ient 
29 Press SYSTEM, SERVICE MENU, TESTS, 33, x1, EXECUTE TEST. 
30 When the analyzer finishes the test, repeat  Step 25 through Step 28.