4-48
Performance Tests
13. Test Port Receiver Magnitude Dynamic Accuracy
Figure4-29. Test Port 1 Input Impedance Test Setup
28 Press MARKER to activate the analyzers marker 1. Use the front panel knob to
locate the maximum value of the data trace for each of the frequency ranges
listed in the Performance Test Record.
29 Write the maximum values on the Performance Test Record.
In Case of
Difficulty
1Suspect the A10 digital IF board assembly if the analyzer fails both test port
tests.
2Refer to the Agilent8753D Service Guide for more troubleshooting
information.
13. Test Port Receiver Magnitude Dynamic Accuracy
Perform this procedure to measure the magnitude dynamic accuracy for each
input test port.