4-48
Performance Tests
13. Test Port Receiver Magnitude Dynamic Accuracy
Figure4-29. Test Port 1 Input Impedance Test Setup 
28 Press MARKER to activate the analyzer’s marker 1. Use the front panel knob to 
locate the maximum value of the data trace for each of the frequency ranges 
listed in the “Performance Test Record.” 
29 Write the maximum values on the “Performance Test Record.” 
In Case of 
Difficulty 
1Suspect the A10 digital IF board assembly if the analyzer fails both test port 
tests. 
2Refer to the Agilent8753D Service Guide for more troubleshooting 
information. 
13. Test Port Receiver Magnitude Dynamic AccuracyPerform this procedure to measure the magnitude dynamic accuracy for each 
input test port.