6-64
Performance Tests – Option 011
16. Source and Receiver Harmonics (Option 002 Only)
13 Press SYSTEM, HARMONIC MEAS, HARMONIC OFF, DISPLAY. 
14 After one sweep, press DATA→MEMORY, DATA/MEM to normalize the trace and 
record the fundamental. 
15 Press SYSTEM, HARMONIC MEAS, HARMONIC THIRD to prepare the analyzer for 
making a 3rd harmonic measurement. 
16 After one sweep, press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH 
MAX. 
17 Write the marker 1 value on the "Performance Test Record." 
Source and Receiver Harmonics 
18 Connect the equipment as shown in Figure 6-33. 
Figure6-33. Source and Receiver Harmonics Test Setup 
19 Press PRESET, MENU, POWER, 6, x1. 
20 Press SYSTEM, HARMONIC MEAS, HARMONIC OFF, STOP, 1.5, G/n.
21 Only for Analyzers with Option 006: Press STOP, 3, G/n.
22 Repeat Step 6 through Step 9 to measure the source and receiver worst case 
2nd harmonic for the receiver input A.
23 Repeat Step 11 through Step 16 to measure the source and receiver worst case 
3nd harmonic for the receiver input A.