6-38
Performance Tests Option 011
11. Receiver Input Impedance
16 Connect the equipment as shown in Figure 6-19.
Figure6-19. Receiver R Input Impedance Test Calibration Test Setup
17 Press PRESET, MEAS, INPUT PORTS, A/B, MENU, NUMBER of POINTS, 1601, x1.
18 Press CAL, CALIBRATE MENU, S11 1-PORT. Repeat Step 4 through Step 6 to perform
a S11 1-port calibration.
19 Connect the equipment as shown in Figure 6-20.
Figure6-20. Receiver R Input Impedance Test Setup