6-9
Performance Tests Option 011
3. Receiver Minimum R Channel Level
31 Press MENU, CW FREQ, 300, k/m. Set the power calibration factor for this CW
frequency.
32 On the Performance Test Record, copy the Path Loss value, previously
calculated for this CW frequency, into the Pass Loss column.
33 Calculate the calibrated source power level using this formula:
Calibrated Power Level = 10 Loss
34 Write the result in the Calibrated Power Level column.
35 Write the power meter reading under the Measured Value column on the
Performance Test Record.
36 Calculate the power level accuracy using this formula:
Calibrated Power Level Measured Value = Power Level Accuracy
37 Write this value on the Performance Test Record.
38 Repeat Step 31 through Step 37 for the other CW frequencies listed on the
Performance Test Record.
In Case of
Difficulty
1Ensure that the power meter and power sensor are operating to specification.
2Inspect the power splitter connectors. Poor match at these connections can
generate power reflections that can cause the analyzer to appear to be out of
specification.
3Inspect the analyzer RF OUT connector for damage.
4If any test fails, perform the RF Output Power Correction Constants (Test
47), located in the Adjustments and Correction Constants chapter of the
Agilent 8753D Option 011 Service Guide.
3. Receiver Minimum R Channel Level
Perform this test to verify the minimum R channel input power level at which
phase lock can be accomplished.