4-3
Performance Tests
1. Test Port Output Frequency Range and Accuracy
WARNING Any servicing, adjustment, maintenance, or repair of this product 
must be performed only by qualified personnel. Repair information 
provided in the In Case of Difficulty sections of this chapter are 
provided for qualified service personnel and are intended to be used 
with the Agilent 8753D Service Guide.
1. Test Port Output Frequency Range and AccuracyPerform this test to verify the frequency accuracy of the Agilent 8702D over 
its entire operating frequency range. 
Required 
Equipment 
Frequency Counter   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilen t5 343A
Cable, 50Ω, Type-N, 24-inch  . . . . . . . . . . . . . . . . . . . . Ag ilent P/N 8120-4781
Adapter, APC-3.5 (f) to Type-N (f) . . . . . . . . .  . . . . . . Agi lent P/N 1250-1745
Adapter, APC-7 to Type-N (f) . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 11524A
Adapter, Type-N (f) to BNC (m) . . . . . . . . . . . . . . . . . Ag ilent P/N 1250-1477
Additional Equipment for Agilent8702D Option 075:
Minimum Loss Pad, 50Ω to 75Ω  . . . . . . . . . . . . . . . . . . . . . . .  Agilent 11852B
Procedure
1Connect the equipment as shown in Figure 4-1. 
Agilent8753D is shown in the figures
The procedures in this chapter are designed to work with Agilent 8702D and 
Agilent8753D instruments. Because of this, many of the figures label the instrument 
being tested as an Agilent8753D. Simply substitute your Agilent 8702D for the 
Agilent8753D.