4-20
Performance Tests
7. Test Port Input Frequency Response
Figure4-10. Setup for Power Meter Calibration on Test Port 1
3Press PRESET, START, 300, k/m.
4Only for Analyzers with Option 006: Press STOP, 3, G/n.
5Press LOCAL, SYSTEM CONTROLLER.
6Press SET ADDRESSES and POWER MTR until the analyzer shows the correct power
meter model.
7Press ADDRESS: P MTR/HPIB. The default power meter GPIB address is 13. Make
sure it is the same as your power meter GPIB address. Otherwise, use the
analyzer front panel keypad to enter the correct GPIB address for your power
meter.