4-12
Performance Tests
5. Minimum R Channel Level
8If the analyzer displays a phase lock error message, continue increasing the test 
port power until phase lock is achieved. 
9Write the test port power, that is displayed on the analyzer, on the 
"Performance Test Record." 
10 Repeat Step 5 through Step 9 for the other CW frequencies listed in the 
"Performance Test Record." 
In Case of 
Difficulty 
WARNING These servicing instructions are for use by qualified personnel only. 
To avoid electrical shock, do not perform any servicing unless you  are 
qualified to do so.
WARNING The opening of covers or removal of parts is likely to expose 
dangerous voltages. Disconnect the instrument from all voltage 
sources while it is being opened.
WARNING The power cord is connected to i nternal capacitors that may remain 
live for five seconds after disconnecting the plug from its power 
supply.
1Check the flexible RF cable (W8, as shown in Figure4-6) between the R 
sampler assembly (A4) and the A11 phase lock assembly. Make sure it is 
connected between A11J1 (PL IF IN) and 1st IF Out.