6-28
Performance Tests – Option 011
9. Receiver Input Crosstalk
(Only for Option 006 Analyzers) R into A Crosstalk from 3 GHz to
4.5 GHz
9Press START, 3, G/n, STOP, 4.5, G/n, NUMBER of GROUPS, 5, x1.
10 When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the “Performance Test Record.”
(Only for Option 006 Analyzers) R into A Crosstalk from 4.5 GHz to
6 GHz
11 Press START, 4.5, G/n, STOP, 6, G/n, NUMBER of GROUPS, 5, x1.
12 When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the “Performance Test Record.”
R into B Crosstalk from 300 kHz to 1 GHz
13 Press START, 300, k/m, STOP, 1, G/n.
14 Press MEAS, INPUT PORTS, B/R, MENU, TRIGGER MENU, NUMBER of GROUPS, 5, x1.
15 When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the “Performance Test Record.”
R into B Crosstalk from 1 GHz to 3 GHz
16 Press START, 1, G/n, STOP, 3, G/n, MENU, TRIGGER MENU, NUMBER of GROUPS, 5, x1.
17 When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the “Performance Test Record.”
(Only for Option 006 Analyzers) R into B Crosstalk from 3 GHz to
4.5 GHz
18 Press START, 3, G/n, STOP, 4.5, G/n, MENU, TRIGGER MENU, NUMBER of GROUPS, 5, x1.
19 When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the “Performance Test Record.”
(Only for Option 006 Analyzers) R into B Crosstalk from 4.5 GHz to
6 GHz
20 Press START, 4.5, G/n, STOP, 6, G/n, MENU, TRIGGER MENU, NUMBER of GROUPS, 5, x1.
21 When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the “Performance Test Record.”