3-7
Manual Verification
Step 3. Device Verification
tude and phase for the four S-parameters).
10 When the Agilent8702D displays, Ver Dev 1 FAIL, proceed to the next step
only if all the tested S-parameter measurements show a PASS status after you
activate error correction in Step 8. Otherwise, refer to In Case of Difficulty on
page 3-9.
11 Connect the 50 dB attenuator (supplied in the verification kit ) as shown in
Figure3-4.
Figure3-4. Connections for the 50 dB Verification Device
12 Press TESTS, 29, x1, EXECUTE TEST, CONTINUE.
Ignore the error message.
13 Press MENU, POWER, 2, x1, CAL, CORRECTION ON.
14 Repeat Step 5 through Step 8 until you have made all the S-parameter
measurements for the verification device.
15 When the Agilent 8702D displays, Ver Dev 2 FAIL, proceed to the next step
only if all S-parameter measurements show a PASS status under the same
condition mentioned in Step 10.
16 Connect the verification mismatch (supplied in the kit) as shown in Figure 3-5.
Be sure port A of the verification mismatch is connected to reference test
port 1.