6-63
Performance Tests Option 011
16. Source and Receiver Harmonics (Option 002 Only)
Figure6-32. Source Harmonics Test Setup
2Press PRESET, MENU, POWER, 20, x1.
3Press START, 16, M/µ, STOP, 1.5, G/n to set the frequency range for searching the
source 2nd harmonic.
4Only for Analyzers with Option 006: Press STOP, 3, G/n.
5Press AVG, IF BW, 10, x1 to set the IF bandwidth to 10 Hz.
6Press MEAS, INPUT PORTS, A.
7After one sweep, press DISPLAY, DATAMEMORY, DATA/MEM to normalize the
trace and record the fundamental.
8Press SYSTEM, HARMONIC MEAS, HARMONIC SECOND to prepare the analyzer for
making a 2nd harmonic measurement.
9After one sweep, press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH
MAX.
10 Write the marker 1 value (which appears on the analyzer display) on the
"Performance Test Record." This is the worst case source 2nd harmonic.
11 Press STOP, 1, G/n to change the stop frequency to 1 GHz.
12 Only for Analyzers with Option 006: Press STOP, 2, G/n.