4-22
Performance Tests
7. Test Port Input Frequency Response
aFrom the sensor calibration factor entries menu, press ADD.
bPress FREQUENCY, 100, k/m. If you make an entry error, press and re-enter
the correct value again.
cPress CAL FACTOR, 96.4, x1.
dPress DONE to terminate the first calibration factor data point entry.
eTo enter the second cal factor data point, press ADD.
fPress FREQUENCY, 300, k/m.
gPress CAL FACTOR, 98.4, x1.
hTo terminate the second calibration factor data point entry, press DONE.
iPress SEGMENT and use the front panel knob to scroll through the sensor cal-
ibration factors table. Check to be sure a ll values are entered correctly. If
you spot an error, use the front panel knob to point to the data point you
want to modify and press EDIT.
13 Press the appropriate softkeys to create a power sensor calibration factors
table.
14 Press DONE to exit the sensor calibration factor entries menu.
15 Press RETURN, ONE SWEEP, TAKE CAL SWEEP to start the power meter calibration.
Wait until the analyzer finishes the sweep, then continue with this procedure.
NOTE
The analyzer displays the PC annotation, indicating the power meter calibration is done
and the error correction is active.