3-3
Manual Verification
Step 1. Initialization
Figure3-1. System Verification Test Setup
2While the equipment is warming up, review the connector care information in
the Agilent 8702D Users Guide.
3Set up the system to print the verification data:
aPress LOCAL, SYSTEM CONTROLLER.
bPress SET ADDRESSES, PRINTER PORT.
cPress the softkey that corresponds to your printer port.
dIf you have difficulty with the printer:
If the interface on your printer is GPIB, verify that the printer address is
set to 1.
If the interface on your printer is serial or parallel, be sure that you cor-
rectly selected the printer port and printer type (refer to the
Agilent 8702D Users Guide for more information on how to do these
tasks).
4Insert the verification kit disk into the Agilent8702D disk drive.
5Press PRESET, SAVE/RECALL, SELECT DISK, INTERNAL DISK.
6Press SYSTEM, SERVICE MENU, TESTS, SYS VER TESTS, EXECUTE TEST.
7When the Agilent8702D displays, Sys Ver Init DONE, the test is complete.