3-9
Manual Verification
In Case of Difficulty
Figure3-6. Mismatch Device Verification Setup 
22 Press TESTS, 31, x1, EXECUTE TEST, CONTINUE. 
Ignore the displayed error message. 
23 Press MENU, POWER, –2, x1, CAL, CORRECTION ON. 
24 Repeat Step 5 through Step 8 until you have made all the S-parameter 
measurements for the verification device. 
25 The test is complete when the Agilent 8702D displ ays, Ver Def 4 FAIL, and 
all S-parameter measurements show a PASS status under the same condition 
mentioned in Step 10. 
In Case of Difficulty 
1Inspect all connections. Do not disconnect the cables from the Agilent8702D 
test ports. Doing so will invalidate the calibration done earlier. 
2Press PRESET, SAVE/RECALL, SELECT DISK, INTERNAL MEMORY, RETURN. Using the 
front panel knob, highlight the full 2-port cal done earlier and press RECALL 
STATE. 
3Repeat “Step 3. Device Verification” on page 3-5.
4If the test still fails, do the following to check the calibration: