4-9
Performance Tests
4. Test Port Output Power Range and Linearity
service manual for more information on how to do this task.
2On the analyzer, press PRESET, MENU, CW FREQ, 300, k/m. Set the power meter
calibration factor for this CW frequency.
3Connect the equipment as shown in Figure 4-4.
Figure4-4. Test Port Output Power Range and Accuracy Test Setup
4On the Agilent438A, press REL. This sets the curr ent power level for relative
power measurement.
5On the analyzer, press MENU, POWER, PWR RANGE MAN.