6-44
Performance Tests Option 011
12. Receiver Magnitude Dynamic Accuracy
aPress CAL, CALIBRATE MENU, RESPONSE, THRU.
bAfter the analyzer finishes the measurement, press DONE: RESPONSE.
cPress SAVE/RECALL, SAVE STATE.
22 Set the Agilent 8496A to 0 dB.
23 On the Agilent 8702D Option 011, press MENU, TRIGGER MENU, SINGLE.
24 Write the mean value in the "Receiver Channel B" column of the "Performance
Test Record."
25 Write the Agilent 8902A readout in the "8902A" column of the "Performance
Test Record."
26 Calculate the difference between the "Receiver Channel B" and the "8902A".
27 Take the absolute value of the calculated difference and enter the result in the
"Magnitude Value" column of the "Performance Test Record."
28 Set the Agilent 8496A to the other attenuation settings listed in the
"Performance Test Record."
29 On the Agilent 8702D Option 011, press SINGLE. Repeat Step 23 through Step
27.
Channel R Magnitude Dynamic Accuracy
30 Connect the equipment as shown in Figure 6-24.