4-56
Performance Tests
14. Test Port Receiver Phase Dynamic Accuracy
Figure4-33. Dynamic Accuracy (Phase)
Procedure Test Port 2 Phase Dynamic Accuracy
1Refer to the “Test Port Receiver Magnitude Dynamic Accuracy” portion of the
“Performance Test Record,” for the values in the “Calculated Value” column.
2Copy the values in the “Magnitude Value” column to the “Calculated Value”
column of the “Test Port Receiver Phase Dynamic Accuracy” portion of the
“Performance Test Record.”
3For every test port input power level, calculate “A” using the following formula:
4Write the result in the “A” column of the “Performance Test Record.”
5For every test port input power level, calculate “B” using the following formula:
B = 1 – A
6Write the result in the “B” column of the “Performance Test Record.”
A10 CalculatedValue
20
---------------------------------------------
=