6-45
Performance Tests Option 011
12. Receiver Magnitude Dynamic Accuracy
Figure6-24. Channel R Magnitude Dynamic Accuracy Test Setup
31 Set the Agilent 8496A to 20 dB.
32 On the Agilent 8702D Option 011:
aPress CONTINUOUS, MEAS, INPUT PORTS, R, MENU, POWER.
bUsing the analyzer front panel knob, adjust the test port power for a reading
of 0.000 dB on the Agilent 8902A.
cWait for the reading to stabilize.
33 On the Agilent 8902A, press the blue shift key and SET REF.
34 On the Agilent 8702D Option 011:
aPress CAL, CALIBRATE MENU, RESPONSE, THRU.
bAfter the analyzer finishes the measurement, press DONE: RESPONSE.
cPress SAVE/RECALL, SAVE STATE.
35 Set the Agilent 8496A to 0 dB.