4-69
Performance Tests
18. Test Port Output/Input Harmonics (Option 002 Analyzers with Option 006 Only)
27 Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH: MAX. 
28 Write the marker 1 value (which appears on the analyzer display) on the 
“Performance Test Record.” This is the worst case port 1 input (receiver 
channel A) 2nd harmonic. 
Port 1 Input Worst Case 3rd Harmonic 
29 Press STOP, 2, G/n to change the stop frequency for measuring the receiver 3rd 
harmonic. 
30 Press SYSTEM, HARMONIC MEAS, HARMONIC OFF. 
31 After one sweep, press DISPLAY, DATA→MEMORY, DATA/MEM to normalize the 
trace. 
32 Press SCALE REF, AUTO SCALE, SCALE/DIV, 1, x1 to get a better viewing of the trace. 
33 Press SYSTEM, HARMONIC MEAS, HARMONIC THIRD. 
34 After one sweep, press SCALE REF, AUTO SCALE. 
35 Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH:MAX. 
36 Write the marker 1 value on the “Performance Test Record.” 
37 Press SYSTEM, HARMONIC MEAS, HARMONIC OFF. 
Port 2 Input Worst Case 2nd Harmonic 
38 Press STOP, 3, G/n to set the stop frequency for measuring the 2nd harmonic. 
39 Press MEAS, Trans: E/E S21 FWD, INPUT PORTS, B. 
40 After one sweep, press DISPLAY, DATA→MEMORY, DATA/MEM to normalize the 
trace. 
41 Press SYSTEM, HARMONIC MEAS, HARMONIC SECOND. 
42 After one sweep, press SCALE REF, AUTO SCALE to get a better viewing of the 
trace. 
43 Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH MAX. 
44 Write the marker 1 value (which appears on the analyzer display) on the 
“Performance Test Record.” This is the worst case port 2 input (receiver 
channel B) 2nd harmonic. 
Port 2 Input Worst Case 3rd Harmonic 
45 Press STOP, 2, G/n to change the stop frequency for measuring the receiver 3rd