5-21
Automated Verification – Option 011
In Case of Difficulty
cPress RECALL STATE. 
dRepeat “Step 3. Device Verification” on page 5-16. 
3If the analyzer still fails the test, check the measurement calibration as follows: 
aPress PRESET. 
bRecall the calibration by pressing SAVE/RECALL, SELECT DISK, INTERNAL DISK, RE-
TURN. 
cUse the front panel knob to highlight the calibration you want to recall and 
press RECALL STATE. 
dConnect the short to reference test port 1. 
ePress SCALE REF, SCALE/DIV, .01, x1. 
fCheck that the trace response is 0.00 ±0.05 dB. 
gDisconnect the short and connect it to reference test port  2. 
hPress MEAS, Refl: E S11 FWD. 
iCheck that the trace response is 0.00 ± 0.05 dB. 
jIf any of the trace responses are out of the specified limits, repeat the “Mea-
surement Calibration” and “Device Verification” procedures. 
4Refer to the Agilent8753 D Option 011 Service Guide for more 
troubleshooting information.