4-34
Performance Tests
9. Calibration Coefficients
9. Calibration Coefficients
Perform this procedure to verify the analyzer uncorrected test port character-
istics.
Required
Equipment
For 50 Analyzers
Calibration Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 85031B
Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent P/N 8120- 4779
For 75 Analyzers
Calibration Kit, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 85036B
Cable, Type-N, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4781
Procedure
First Full 2-Port Calibration
1Connect the equipment as shown in Figure 4-20.
Figure4-20. First Full 2-Port Calibration Test Setup
NOTE
The crosstalk calibration coefficients are omitted in this procedure. They are covered in
the Test Port Crosstalk performance test.