6-34
Performance Tests Option 011
10. Receiver Trace Noise
statistic feature.
5Write the s.dev (standard deviation) value, which appears on the analyzer
display, on the "Performance Test Record."
6Press MEAS, INPUT PORTS, B/R. Write the s.dev (standard deviation) value on the
"Performance Test Record."
7Press A/B. Write the s.dev (standard deviation) value on the "Performance Test
Record."
8Press FORMAT, PHASE. Write the s.dev (standard deviation) value on the
"Performance Test Record."
9Press MEAS, INPUT PORTS, B/R, FORMAT, PHASE. Write the s.dev (standard
deviation) value on the "Performance Test Record."
10 Press MEAS, INPUT PORTS, A/R, FORMAT, PHASE. Write the s.dev (standard
deviation) value on the "Performance Test Record."
In Case of
Difficulty
WARNING These servicing instructions are for use by qualified personnel only.
To avoid electrical shock, do not perform any servicing unless you are
qualified to do so.
WARNING The opening of covers or removal of parts is likely to expose
dangerous voltages. Disconnect the instrument from all voltage
sources while it is being opened.
WARNING The power cord is connected to i nternal capacitors that may remain
live for five seconds after disconnecting the plug from its power
supply.
1Verify that the 50 Type-N RF cables are in good condition. If possible, try to
rerun this test with another set of RF cables.
2Suspect the A10 Digital IF board assembly if the analyzer still fails the test.