6-48
Performance Tests Option 011
14. Receiver Magnitude Compression
4Write the result in the "A" column of the "Performance Test Record."
5For every receiver input power level, calculate "B" using the following formula:
B = 1 A
6Write the result in the "B" column of the " Performance Test Record."
7For every receiver input power level, calculate the receiver phase dynamic
accuracy using the following formula:
Phase Dynamic Accuracy = SIN 1 (B)
8Write the value in the "Calculated Value" column of the "Performance Test
Record."
Channel B Phase Dynamic Accuracy
9Repeat Step 3 through Step 8.
Channel R Phase Dynamic Accuracy
10 Repeat Step 3 through Step 8.
14. Receiver Magnitude Compression
Perform this test to verify the compression/expansion magnitude levels of the
analyzers receiver samplers.
Required
Equipment
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8482A
Attenuator, 3 dB. . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8491A Option 003
Attenuator, 10 dB. . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8491A Option 010
Attenuator, 20 dB. . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8491A Option 020
Cable (2), 50, Type-N, 24-inch . . . . . . . . . . . . . . . . . .Agilent P/N 8120-4781
2-way Power Splitter (2), 50. . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 11667A
Adapter, Type-N (m) to Type-N (m) . . . . . . . . . . . . . .Agilent P/N 1250-1475
A10 CalculatedValue