5-16
Automated Verification – Option 011
Step 3. Device Verification
10 When the analyzer finishes measuring the standard, press STANDARDS DONE.
11 Connect the test port cables as shown Figure5-10.
Figure5-10. Transmission Calibration Setup
12 Press TRANSMISSION, FWD TRANS THRU.
13 When the analyzer finishes the measurement , press FWD MATCH THRU.
14 When the analyzer displays, PRESS ’DONE’ IF FINISHED WITH STD(s),
press STANDARDS DONE, DONE 2-PORT CAL.
15 Press SAVE/RECALL, SELECT DISK, INTERNAL MEMORY, RETURN, SAVE STATE to save
the calibration into the analyzer’s internal memory.
16 When the analyzer finishes storing the measurement cal ibration, press SELECT
DISK, INTERNAL DISK.
Step 3. Device Verification
1Press SYSTEM, SERVICE MENU, TESTS, 28, x1, EXECUTE TEST.
2At the prompt, connect the 20 dB attenuator (supplied in the verification kit)