6-68
Performance Tests Option 011
17. Receiver Magnitude Frequency Response (Option 002 Only )
Test Record.
8Press B. Write the marker 1 reading, which appears on the analyzer display, in
the Input B Value column on the Performance Test Record.
9On the external source, press 48, MHz. Record the power meter reading on t he
Performance Test Record.
10 On the analyzer, press SYSTEM, HARMONIC MEAS, THIRD. Write the marker 1
reading in the Input B Value column on the Performance Test Record.
11 Press MEAS, INPUT PORTS, A. Write the marker 1 reading, which appears on the
analyzer display, in the Input A Value column on the Performance Test
Record.
12 On the analyzer, press SYSTEM, HARMONIC MEAS, OFF, MENU, CW FREQ, 31, M/µ to
set the analyzer to the next fundamental frequency.
13 On the external source, press 62, MHz. Record the power meter reading on t he
Performance Test Record.
14 On the analyzer, press SYSTEM, HARMONIC MEAS, SECOND. Write the marker 1
reading in the Input A Value column on the Performance Test Recor d.
15 Press MEAS, INPUT PORTS, B. Write the marker 1 reading, which appears on the
analyzer display, in the Input B Value column on the Performance Test
Record.
16 On the external source, press 93, MHz. Record the power meter reading on t he
Performance Test Record.
17 On the analyzer, press SYSTEM, HARMONIC MEAS, THIRD. Write the marker 1
reading in the Input B Value column on the Performance Test Record.
18 Press MEAS, INPUT PORTS, A. Write the marker 1 reading, which appears on the
analyzer display, in the Input A Value column on the Performance Test
Record.
19 Repeat Step 12 through Step 18 for the CW frequencies listed on the
Performance Test Record.
20 For each measurement, calculate the difference between the measurement
and the power meter level.
21 Write the calculation results in the Performance Test Record.
22 Compare the results with the specification .