5-20
Automated Verification – Option 011
In Case of Difficulty
8When the analyzer finishes all the measurements, reconnect the mismatch
verification device, as shown in Figure 5-14.
Figure5-14. Mismatch Device Verification Setup
9Press RETURN, TESTS, 31, x1, EXECUTE TEST, CONTINUE.
10 You have completed the system verification procedure when the analyzer
displays, Ver Def 4 PASS.
In Case of Difficulty
1Inspect all connections. Do not disconnect the cables from the analyzer test
ports. Doing so will invalidate the calibration that you have done earlier.
2Recall the measurement calibration and repeat the procedure:
aPress PRESET, SAVE/RECALL, SELECT DISK, INTERNAL DISK, RETURN.
bUsing the front panel knob, highlight the title of the one-path 2-port calibra-
tion that you have done earlier.