5-4
Automated Verification Option 011
Agilent8702D Option 011 and Agilent 85046A/47A System Verification
Agilent 8702D Option 011 and Agilent 85046A/47A System Verification
In order to run this procedure, the following equipment is required:
S-Parameter test set (for analyzers without Option 006) . . . .Agilent 85046A
S-Parameter test set (for analyzers with Option 006) . . . . . . .Agilent 85047A
Calibration Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 85031B
Verification Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 85029B
RF Cable Set, 7 mm, 50. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 11857D
Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP ThinkJet/DeskJet/LaserJet
Step 1. Initialization
1Connect the equipment as shown in Figure 5-1. Let the Ag ilent 8702D
Option 011 warm up for one hour.
NOTE
Do not proceed any further if your system has an Agilent 85044A Transmission/Reflec-
tion test set. Instead, go to Agilent8702D Option 011 and Agilent 85044A System Veri-
fication on page5-13.