4-24
Performance Tests
7. Test Port Input Frequency Response
Power Meter Calibration on Port 2 from 300 kHz to 3 GHz
22 Connect the equipment as shown Figure 4-12.
Figure4-12. Setup for Power Meter Calibration on Test Port 2
23 Press MEAS, INPUT PORTS, TESTPORT 2.
24 Press CAL, PWRMTR CAL, ONE SWEEP, TAKE CAL SWEEP to start the power meter
calibration for test port 2.