
4-10
Performance Tests
5. Minimum R Channel Level
6Write the power meter reading in the "Results Measured" column on the 
"Performance Test Record." 
7Calculate the difference between the analyzer test port power (which appears 
on the analyzer’s display) and the power meter reading. Write the result in the 
"Power Level Linearity" column on the "Performance Test Record." 
8Repeat Step 5 through Step 7 for the other power levels listed in the 
"Performance Test Record." 
9Press CW FREQ, 3, G/n. 
10 Set the power meter calibration factor for this CW frequency and press REL to 
set the reference at this new frequency. 
11 Press MENU, POWER, –15, x1. 
12 Write the power meter reading in the "Results Measured" column on the 
"Performance Test Record." 
13 Calculate the difference between the analyzer test port power and the power 
meter reading. Write the result in the "Power Level Linearity" column of the 
"Performance Test Record." 
14 Repeat Step 11 through Step 13 for the other power levels listed in the 
"Performance Test Record." 
In Case of 
Difficulty 
1Ensure that the power meter and power sensor(s) are operating to 
specifications. Be sure you set the power meter calibration factor for the CW 
frequency that you are testing. 
2Verify that there is power coming out of the analyzer’s test port 1. Be sure you 
did not accidently switch off the analyzer’s internal source. If you did so, press 
MENU, POWER, SOURCE PWR ON. 
3Repeat this performance test. 
5. Minimum R Channel LevelPerform this test to determine the minimum R channel input power level at 
which phase lock can be accomplished.