5-7
Automated Verification Option 011
Step 2. Measurement Calibration
7Press FORWARD:SHORT.
8When the analyzer displays the message CONNECT STD THEN PRESS KEY
TO MEASURE, connect the 50 ohm termination (supplied in the calibration kit)
to reference test port 1, as shown in Figure5-2.
9Press FORWARD:LOAD.
10 When the analyzer finishes measuring the sta ndard, connect the open end of
the open/short combination to reference test port 2, as shown in Figure5-2.
11 Press REVERSE:OPEN.
12 When the analyzer displays the message CONNECT STD THEN PRESS KEY
TO MEASURE, connect the short end of the open/short combination to
reference port 2 as shown in Figure 5-2.
13 Press REVERSE:SHORT.
14 When the analyzer finishes measuring the standard, connect the 50 ohm
termination to reference port 2, as shown in Figure5-2.
15 Press REVERSE:LOAD.
16 When the analyzer finishes measuring the standard, press STANDARDS DONE.
17 Connect the test port cables as shown Figure5-3.
Figure5-3. Transmission Calibration Setup