4-41
Performance Tests
10. System Trace Noise (Only for Analyzers without Option 006)
2Press PRESET, MENU, POWER, 5, x1.
3Press RETURN, CW FREQ, 3, G/n, NUMBER of POINTS, 1601, x1.
4Press MARKER FCTN, MARKER MODE MENU, STATS ON to activate the instruments
statistic feature.
System Trace Noise for A/R Magnitude
5Press MEAS, Trans: E/E S12 REV.
6Press MENU, TRIGGER MENU, NUMBER of GROUPS, 5, x1.
7When the analyzer displays the Hld annotation, press SCALE REF, AUTO SCALE.
8Write the s.dev (standard deviation) value, which appears on the analyzer
display, on the Performance Test Record.
System Trace Noise for A/R Phase
9Press FORMAT, PHASE.
10 Press MENU, TRIGGER MENU, NUMBER of GROUPS, 5, x1.
11 When the analyzer finishes the number of sweeps, press SCALE REF, AUTO SCALE.
12 Write the s.dev (standard deviation) value on the Performance Test Record.
System Trace Noise for B/R Magnitude
13 Press MEAS, Trans: E/E S21 FWD.
14 Press MENU, TRIGGER MENU, NUMBER of GROUPS, 5, x1.
15 When the analyzer finishes the number of sweeps, press SCALE REF, AUTO SCALE.
16 Write the s.dev (standard deviation) value on the Performance Test Record.
System Trace Noise for B/R Phase
17 Press FORMAT, PHASE.
18 Press MENU, TRIGGER MENU, NUMBER of GROUPS, 5, x1.
19 When the analyzer finishes the number of sweeps, press SCALE REF, AUTO SCALE.
20 Write the s.dev (standard deviation) value on the Performance Test Record.